Mechanical strength of poly(methyl methacrylate) cement-human bone interfaces

R. Kusleika, Samuel I Stupp

Research output: Contribution to journalArticle

21 Citations (Scopus)

Abstract

A device was constructed to test the interfacial strength of PMMA-based bone cement and human cancellous bone under pure tension. Two types of tissue were used in the investigation: formalin-fixed vertebral bone as an in vitro model for weak cancellous bone, and freshly removed metatarsal bone. Tissue-cement joints were allowed to solidify under two different pressures (0.11 and 0.47 MPa), and cement placement time on tissue surfaces was also controlled as a variable. The higher curing pressure only seemed to enhance the strength of interfaces formed with mechanically weak fixed bone but had no significant effect for joints formed with the stronger, freshly extracted tissue. Cement placement time did not have a discernible effect on interfacial strength regardless of the tissue used or the pressure applied during setting. An analysis of fracture morphology by optical microscopy revealed largely cement cohesive failure in some cases and bone or mixed fractures in others. Joints exhibiting mainly cement fracture had the highest interfacial tensile strengths (in the order of 7.5 MPa). Once measured values of tissue porosity were taken into account, the observed joint strength correlated well with cement tensile strength. Based on experimental findings, better stress-dissipating qualities and higher tensile strength are suggested as two important necessary improvements of bone cements based on poly(methyl methacrylate).

Original languageEnglish
Pages (from-to)441-458
Number of pages18
JournalJournal of Biomedical Materials Research
Volume17
Issue number3
Publication statusPublished - May 1983

Fingerprint

Bone cement
Polymethyl Methacrylate
Polymethyl methacrylates
Strength of materials
Bone
Tissue
Cements
Bone Cements
Tensile strength
Formaldehyde
Optical microscopy
Curing
Porosity

ASJC Scopus subject areas

  • Biomedical Engineering
  • Biomaterials

Cite this

Mechanical strength of poly(methyl methacrylate) cement-human bone interfaces. / Kusleika, R.; Stupp, Samuel I.

In: Journal of Biomedical Materials Research, Vol. 17, No. 3, 05.1983, p. 441-458.

Research output: Contribution to journalArticle

@article{df4a413023764696bba5f0262a2147ea,
title = "Mechanical strength of poly(methyl methacrylate) cement-human bone interfaces",
abstract = "A device was constructed to test the interfacial strength of PMMA-based bone cement and human cancellous bone under pure tension. Two types of tissue were used in the investigation: formalin-fixed vertebral bone as an in vitro model for weak cancellous bone, and freshly removed metatarsal bone. Tissue-cement joints were allowed to solidify under two different pressures (0.11 and 0.47 MPa), and cement placement time on tissue surfaces was also controlled as a variable. The higher curing pressure only seemed to enhance the strength of interfaces formed with mechanically weak fixed bone but had no significant effect for joints formed with the stronger, freshly extracted tissue. Cement placement time did not have a discernible effect on interfacial strength regardless of the tissue used or the pressure applied during setting. An analysis of fracture morphology by optical microscopy revealed largely cement cohesive failure in some cases and bone or mixed fractures in others. Joints exhibiting mainly cement fracture had the highest interfacial tensile strengths (in the order of 7.5 MPa). Once measured values of tissue porosity were taken into account, the observed joint strength correlated well with cement tensile strength. Based on experimental findings, better stress-dissipating qualities and higher tensile strength are suggested as two important necessary improvements of bone cements based on poly(methyl methacrylate).",
author = "R. Kusleika and Stupp, {Samuel I}",
year = "1983",
month = "5",
language = "English",
volume = "17",
pages = "441--458",
journal = "Journal of Biomedical Materials Research - Part A",
issn = "0021-9304",
publisher = "John Wiley and Sons Inc.",
number = "3",

}

TY - JOUR

T1 - Mechanical strength of poly(methyl methacrylate) cement-human bone interfaces

AU - Kusleika, R.

AU - Stupp, Samuel I

PY - 1983/5

Y1 - 1983/5

N2 - A device was constructed to test the interfacial strength of PMMA-based bone cement and human cancellous bone under pure tension. Two types of tissue were used in the investigation: formalin-fixed vertebral bone as an in vitro model for weak cancellous bone, and freshly removed metatarsal bone. Tissue-cement joints were allowed to solidify under two different pressures (0.11 and 0.47 MPa), and cement placement time on tissue surfaces was also controlled as a variable. The higher curing pressure only seemed to enhance the strength of interfaces formed with mechanically weak fixed bone but had no significant effect for joints formed with the stronger, freshly extracted tissue. Cement placement time did not have a discernible effect on interfacial strength regardless of the tissue used or the pressure applied during setting. An analysis of fracture morphology by optical microscopy revealed largely cement cohesive failure in some cases and bone or mixed fractures in others. Joints exhibiting mainly cement fracture had the highest interfacial tensile strengths (in the order of 7.5 MPa). Once measured values of tissue porosity were taken into account, the observed joint strength correlated well with cement tensile strength. Based on experimental findings, better stress-dissipating qualities and higher tensile strength are suggested as two important necessary improvements of bone cements based on poly(methyl methacrylate).

AB - A device was constructed to test the interfacial strength of PMMA-based bone cement and human cancellous bone under pure tension. Two types of tissue were used in the investigation: formalin-fixed vertebral bone as an in vitro model for weak cancellous bone, and freshly removed metatarsal bone. Tissue-cement joints were allowed to solidify under two different pressures (0.11 and 0.47 MPa), and cement placement time on tissue surfaces was also controlled as a variable. The higher curing pressure only seemed to enhance the strength of interfaces formed with mechanically weak fixed bone but had no significant effect for joints formed with the stronger, freshly extracted tissue. Cement placement time did not have a discernible effect on interfacial strength regardless of the tissue used or the pressure applied during setting. An analysis of fracture morphology by optical microscopy revealed largely cement cohesive failure in some cases and bone or mixed fractures in others. Joints exhibiting mainly cement fracture had the highest interfacial tensile strengths (in the order of 7.5 MPa). Once measured values of tissue porosity were taken into account, the observed joint strength correlated well with cement tensile strength. Based on experimental findings, better stress-dissipating qualities and higher tensile strength are suggested as two important necessary improvements of bone cements based on poly(methyl methacrylate).

UR - http://www.scopus.com/inward/record.url?scp=0020751412&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020751412&partnerID=8YFLogxK

M3 - Article

VL - 17

SP - 441

EP - 458

JO - Journal of Biomedical Materials Research - Part A

JF - Journal of Biomedical Materials Research - Part A

SN - 0021-9304

IS - 3